New Features of the
REMEDI 1000
The system can be operated in two modes, the 'MONITORED" mode
which complies with AS 2896 enables the panel to distinguish between
the usual switched functions while at the same time the panel monitors
the integrity of the wiring where "SYSTEM SHORT CIRCUIT" and "SYSTEM
OPEN CIRCUIT" are able to be detected. In its alternate mode (switched
only) the panel can be used to monitor the "ON" and "OFF" functions
only.
The Facia panel is manufactured from Polycarbonate, the surface
finish allows easy cleaning with all alarm controls and instructions
under the surface. The fluorescent display will accept messages
with up to 40 characters and the memory holds 24 individual messages.
The panel Facia also has an infrared receiver window, this window
allows the programming of the panel while IN operation OR prior
to commissioning, messages may be altered to suit the particular
hospital at any time, all that is needed is a Laptop computer and
an infrared transmitter which communicates through the panel Facia.
When the panel is in "programming mode" all system operating parameters
can be modified.
As part of the Remedi
1000 system the Alarm Panel also has a matching BMS interface,
this unit is installed in parallel with the Alarm Panel and incorporates
24 clean contacts with the option of either Normally Open or Closed
terminals. When the Remedi
1000 Alarm Panel is in "Test" mode it will also operate
the BMS Interface contacts.